Verificación del desempeño analítico de marcadores bioquímicos de función renal implementados en una plataforma automatizada en el servicio de laboratorio clínico de un hospital en Lima, Perú

dc.contributor.advisorFernández Olivera, César Augustoes_ES
dc.contributor.advisorZavaleta Martínez-Vargas, Alfonsoes_ES
dc.contributor.authorDuran Dávila, Carmen Alejandraes_ES
dc.contributor.authorMogollón Reyes, Silvana del Rosarioes_ES
dc.contributor.authorPulache Mauricio, Álvaro Renzoes_ES
dc.date.accessioned2022-06-24T22:07:41Z
dc.date.available2022-06-24T22:07:41Z
dc.date.issued2021
dc.description.abstractAntecedentes: La enfermedad renal crónica afecta la vida de 1 en cada 10 personas a nivel mundial y su incidencia va aumentando, por lo que es necesario que el laboratorio clínico emita resultados confiables para el diagnóstico y seguimiento de pacientes con enfermedades renales. Para lograr esto se debe emplear metodologías analíticas que cumplan las especificaciones técnicas de su fabricante. Objetivos: Verificar el desempeño analítico de las metodologías analíticas de los marcadores de función renal (creatinina, nitrógeno ureico/urea, ácido úrico, proteínas en orina y microalbúmina) implementados en el analizador Dimension RxL Max del Centro Médico Naval, aplicando el protocolo EP 15 A3 del CLSI y comparar el Error total de las metodologías con el Error Total Máximo permitido (ETp) según CLIA y/o Variabilidad Biológica. Materiales y métodos: Se realizó un estudio Descriptivo transversal. Las mediciones de los analitos se realizaron en dos niveles de decisión clínica de control, según la guía del CLSI. Los datos se analizaron en el software estadístico Analyse- it y el programa Microsoft Excel. Resultados: La precisión de repetibilidad y la precisión intralaboratorial verificaron correctamente en todas las metodologías utilizadas. Sin embargo, para veracidad las metodologías de Nitrógeno ureico y Proteínas en orina no verificaron. Solo el error total de ambos métodos superó el ETp según CLIA. Conclusiones: Se puede emplear las metodologías para la detección de creatinina, ácido úrico en sangre y orina y microalbúmina en orina. Se deberá contactar con el fabricante del método de nitrógeno ureico y proteínas en orina para la corrección(es) de error(es) y realizar una nueva verificación del desempeño analítico.es_ES
dc.description.abstractBackground: Chronic kidney disease affects the life of 1 in 10 people worldwide and its incidence is increasing. This is why the clinical laboratory must issue reliable results for the diagnosis and follow-up of patients with renal diseases. This is achieved by using analytical methodologies that meet the technical specifications provided by their manufacturer. Objectives: Verify the analytical performance of the markers for renal function (creatinine, urea nitrogen, uric acid, urine proteins and microalbumin) implemented in the Dimension RXL Max automated system of the Naval Medical Center applying the CLSI EP 15 A3 guidelines. Then, compare the Total Errors of the methods with the Maximum allowable error (ETp) by CLIA and Biological Variability. Methods and Materials: A cross-sectional descriptive study was carried out. All runs were made according to EP 15 A3 guidelines for 2 levels of clinical decision reference materials. The data was analyzed using the statistical software Analyse-It and Microsoft Excel. Results: Repeatability and Intralaboratory precision were verified in all methods. However, the methods for urea and urine protein at normal level didn’t verify in trueness. Also, the total errors of both of these methods surpassed the ETp specified by CLIA. Conclusions: The methods for creatinine, uric acid and microalbumin can continue to be used. However, the manufacturer of the urea and urine protein methods should be contacted to correct the errors and then a new analytical performance verification should be made.es_ES
dc.formatapplication/pdfes_ES
dc.identifier.other66891es_ES
dc.identifier.urihttps://hdl.handle.net/20.500.12866/11831
dc.language.isospaes_ES
dc.publisherUniversidad Peruana Cayetano Herediaes_ES
dc.publisher.countryPEes_ES
dc.rightshttps://purl.org/coar/access_right/c_abf2es_ES
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/4.0/deed.eses_ES
dc.subjectTécnicas de Química Analíticaes_ES
dc.subjectControl de Calidades_ES
dc.subjectSesgoes_ES
dc.subjectReproducibilidad de Resultadoses_ES
dc.subjectPruebas de Función Renales_ES
dc.subject.ocdehttps://purl.org/pe-repo/ocde/ford#1.04.07es_ES
dc.subject.ocdehttps://purl.org/pe-repo/ocde/ford#2.06.02es_ES
dc.subject.ocdehttps://purl.org/pe-repo/ocde/ford#3.02.20es_ES
dc.titleVerificación del desempeño analítico de marcadores bioquímicos de función renal implementados en una plataforma automatizada en el servicio de laboratorio clínico de un hospital en Lima, Perúes_ES
dc.title.alternativeVerification of the analytical performance of biochemical markers of renal function implemented in an automated platform in a clinical laboratory of a hospital in Lima, Peruen_US
dc.typehttp://purl.org/coar/resource_type/c_7a1fes_ES
dspace.entity.typePublication
renati.advisor.dni41907821
renati.advisor.dni08425672
renati.advisor.orcidhttps://orcid.org/0000-0002-8039-9589es_ES
renati.author.dni47210543
renati.author.dni73692912
renati.author.dni71336025
renati.discipline915016es_ES
renati.jurorFigueroa Tataje, Jaime Josées_ES
renati.jurorFlores Barreto, María Emiliaes_ES
renati.jurorSiqueros Huaman, Davides_ES
renati.levelhttps://purl.org/pe-repo/renati/level#tituloProfesionales_ES
renati.typehttps://purl.org/pe-repo/renati/type#tesises_ES
thesis.degree.disciplineTecnología Médica en la Especialidad de Laboratorio Clínicoes_ES
thesis.degree.grantorUniversidad Peruana Cayetano Heredia. Facultad de Medicina Alberto Hurtadoes_ES
thesis.degree.nameLicenciado en Tecnología Médica en la Especialidad de Laboratorio Clínicoes_ES

Archivos

Bloque original

Mostrando 1 - 1 de 1
Cargando...
Miniatura
Nombre:
Verificacion_DuranDavila_Carmen.pdf
Tamaño:
554.57 KB
Formato:
Adobe Portable Document Format
Descripción:

Bloque de licencias

Mostrando 1 - 1 de 1
Cargando...
Miniatura
Nombre:
license.txt
Tamaño:
1.82 KB
Formato:
Item-specific license agreed upon to submission
Descripción: