Accuracy of HemoCue301 portable hemoglobin analyzer for anemia screening in capillary blood from women of reproductive age in a deprived region of Northern Peru: An on-field study.

Mostrar el registro sencillo del ítem

dc.contributor.author Alarcón Yaquetto, Dulce Esperanza
dc.contributor.author Rueda Torres, Lenin Vladimir
dc.contributor.author Bailon Gonzales, Nataly Aracely
dc.contributor.author Barreto, Percy Vílchez
dc.contributor.author Málaga Rodríguez, Germán Javier
dc.coverage.spatial Tumbes, Perú
dc.date.accessioned 2023-12-05T17:47:57Z
dc.date.available 2023-12-05T17:47:57Z
dc.date.issued 2023
dc.identifier.uri https://hdl.handle.net/20.500.12866/14606
dc.description.abstract OBJECTIVE: We aim to assess the accuracy and effectiveness of the HemoCue 301, a point-of-care (POC) device for measuring hemoglobin levels, and detecting anemia among individuals living in Tumbes, a rural, underserved area in Northern Peru. METHODS: Baseline analysis of a clinical trial aimed at assessing the effect of multi-fortified bread (NCT05103709). Adult women with capillary blood HemoCue 301 readings below 12 g/dL were recruited in coastal city of Tumbes, Peru. A total of 306 women took part of the study, venous blood samples were taken and analyzed with an automated hematology analyzer. Serum samples were used to measure ferritin, serum iron and C reactive protein. RESULTS: Capillary blood measured by the Hemocue 301 has a bias of 0.36 ± 0.93 g/dL respect to the automated Hb. More than 50% of women with normal ferritin values were classified as anemics according to the HemoCue 301. Automated Hb cut-off of 10.8 g/dL [AUC 0.82 (0.77-0.88)] had a specificity of 0.817 and a sensitivity 0.711 while with the HemoCue 301 cut-off of 11.1 g/dL [AUC 0.71 (0.62-0.79)] had a specificity of 0.697 and a sensitivity 0.688. The performance of the automated Hb cut-off was significantly better than the HemoCue (p<0.001). CONCLUSION: Caution must be taken when using POC devices, especially with values around the threshold. Cut-off values found in our study could be used as surrogate means when no confirmatory tests are available. Clinical outcomes should be prioritized when diagnosing iron deficiency anemia in women of reproductive age to ensure proper diagnosis. en_US
dc.language.iso eng
dc.publisher PLoS
dc.relation.ispartofseries PLoS One
dc.rights info:eu-repo/semantics/restrictedAccess
dc.rights.uri https://creativecommons.org/licenses/by-nc-nd/4.0/deed.es
dc.subject Ferritin en_US
dc.subject Anemia en_US
dc.subject Blood en_US
dc.subject Hemoglobin en_US
dc.subject Iron deficiency anemia en_US
dc.subject Peru en_US
dc.subject Capillaries en_US
dc.subject Red blood cells en_US
dc.subject.mesh Ferritinas
dc.subject.mesh Anemia
dc.subject.mesh Sangre
dc.subject.mesh Hemoglobinas
dc.subject.mesh Anemia Ferropénica
dc.subject.mesh Perú
dc.subject.mesh Capilares
dc.subject.mesh Eritrocitos
dc.title Accuracy of HemoCue301 portable hemoglobin analyzer for anemia screening in capillary blood from women of reproductive age in a deprived region of Northern Peru: An on-field study. en_US
dc.type info:eu-repo/semantics/article
dc.identifier.doi https://doi.org/10.1371/journal.pone.0293984
dc.subject.ocde https://purl.org/pe-repo/ocde/ford#3.02.06
dc.relation.issn 1932-6203


Ficheros en el ítem

Ficheros Tamaño Formato Ver

No hay ficheros asociados a este ítem.

Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del ítem

info:eu-repo/semantics/restrictedAccess Excepto si se señala otra cosa, la licencia del ítem se describe como info:eu-repo/semantics/restrictedAccess

Buscar en el Repositorio


Listar

Panel de Control

Estadísticas