dc.contributor.author |
Bristow, Claire C. |
|
dc.contributor.author |
Vargas Rivera, Silver Keith |
|
dc.contributor.author |
Ramos Cordova, Lourdes B. |
|
dc.contributor.author |
Qquellon Palacios, Luz Jazmín |
|
dc.contributor.author |
Konda, Kelika A. |
|
dc.contributor.author |
Klausner, Jeffrey D. |
|
dc.date.accessioned |
2019-03-05T15:25:03Z |
|
dc.date.available |
2019-03-05T15:25:03Z |
|
dc.date.issued |
2018 |
|
dc.identifier.uri |
https://hdl.handle.net/20.500.12866/5937 |
|
dc.description.abstract |
We evaluated the Standard Q HIV/Syphilis Combo Test (SD Biosensor, South Korea), a dual rapid test using stored sera (N = 400) in a laboratory setting in Lima, Peru. The sensitivity and specificity for HIV antibody detection was 100.0% (95% CI: 98.2–100.0%) and 99.5% (95% CI: 97.2–100.0%), respectively. For treponemal antibody detection the sensitivity and specificity was 97.5% (95%CI:94.3–99.2%) and 100.0% (95%CI:98.2–100.0%), respectively. |
en_US |
dc.language.iso |
eng |
|
dc.publisher |
Elsevier |
|
dc.relation.ispartofseries |
Diagnostic Microbiology and Infectious Disease |
|
dc.rights |
info:eu-repo/semantics/restrictedAccess |
|
dc.rights.uri |
https://creativecommons.org/licenses/by-nc-nd/4.0/deed.es |
|
dc.subject |
antibody detection |
en_US |
dc.subject |
Article |
en_US |
dc.subject |
cohort analysis |
en_US |
dc.subject |
controlled study |
en_US |
dc.subject |
Diagnosis |
en_US |
dc.subject |
diagnostic accuracy |
en_US |
dc.subject |
diagnostic test accuracy study |
en_US |
dc.subject |
Dual test |
en_US |
dc.subject |
enzyme immunoassay |
en_US |
dc.subject |
female |
en_US |
dc.subject |
HIV |
en_US |
dc.subject |
human |
en_US |
dc.subject |
Human immunodeficiency virus infection |
en_US |
dc.subject |
immunoreactivity |
en_US |
dc.subject |
major clinical study |
en_US |
dc.subject |
Peru |
en_US |
dc.subject |
priority journal |
en_US |
dc.subject |
Rapid test |
en_US |
dc.subject |
sensitivity and specificity |
en_US |
dc.subject |
syphilis |
en_US |
dc.subject |
Treponema pallidum |
en_US |
dc.subject |
Treponema pallidum hemagglutination test |
en_US |
dc.title |
Dual rapid test for HIV and syphilis: A laboratory evaluation of the diagnostic accuracy of the Standard Q HIV/Syphilis Combo Test |
en_US |
dc.type |
info:eu-repo/semantics/article |
|
dc.identifier.doi |
https://doi.org/10.1016/j.diagmicrobio.2018.11.018 |
|
dc.subject.ocde |
https://purl.org/pe-repo/ocde/ford#3.03.08 |
|
dc.subject.ocde |
https://purl.org/pe-repo/ocde/ford#1.06.01 |
|
dc.relation.issn |
1879-0070 |
|